2022 | 1(102) | 127-136

Article title

Wytrzymałościowa analiza multikątowej prowadnicy stosowanej w chirurgii kostnej za pomocą analizy MES

Content

Title variants

EN
Strength analysis of a multi-angle guide used in bone surgery using FEM analyse

Languages of publication

Abstracts

EN
The study carried out a strength analysis of a new construction form of a multi-angle guide enabling the hole to be made at a selected angle (j = ± 15°). The calculations were carried out in order to select the best material variant from which the guide should be made. Geometry verification and selection of tool material was carried out on the basis of results obtained using MES. Polymer (PET, PP, PTFE) and metal (TiAlV, CrNiCo) materials were proposed for the individual elements of the guide.
PL
W pracy przeprowadzono analizę wytrzymałościową nowej postaci konstrukcyjnej multikątowej prowadnicy umożliwiającej wykonanie otworu pod zadanym kątem (j = ± 15°). Obliczenia przeprowadzono w celu wytypowania najkorzystniejszego wariantu materiałowego, z którego prowadnica powinna zostać wykonana. Weryfikację geometrii oraz dobór materiału narzędzia przeprowadzono na podstawie wyników uzyskanych za pomocą MES. Do wykonania poszczególnych elementów prowadnicy zaproponowano zarówno materiały polimerowe (PET, PP, PTFE) jak i metalowe (TiAlV, CrNiCo).

Year

Issue

Pages

127-136

Physical description

Dates

published
2022

Contributors

  • Politechnika Śląska, Zabrze Katedra Biomechatroniki, Wydział Inżynierii Biomedycznej
  • Politechnika Śląska, Zabrze Katedra Biomechatroniki, Wydział Inżynierii Biomedycznej
  • Politechnika Śląska, Zabrze Katedra Biomechatroniki, Wydział Inżynierii Biomedycznej

References

  • BHH MIKROMED, katalog: 2.01 DHS/DCS, 2018.
  • Bishop E., Handel I., Clements D., Ryan J., Sequential drilling and drill angulation reduce the accuracy of drill hole start location in a synthetic bone model, „Vet Rec.” 2019.
  • Brioschi V., Cook J., Arthurs G., Can a surgeon drill accurately at a specified angle?, „Vet Rec Open.” 2016.
  • Giles R. Scuderi, Alfred J., Jr. Tria, Surgical Techniques in Total Knee Arthroplasty, Springer Science & Business Media, 2006, p. 83-105.
  • MEDGAL orthopaedic implants & instruments, katalog: Płytki kostne – system 2.0.
  • Neal S. ElAttrache, Christopher D. Harner, Raffy Mirzayan, Jon K. Sekiya, Surgical Techniques in Sports Medicine, Lippincott Williams & Wilkins, 2007, p. 325-346.
  • Pandey R.K., Panda S.S., Drilling of bone: A comprehensive review, „J Clin Orthop Trauma”, 2013, p. 15-30.
  • Shales CJ., White L., Langley-Hobbs SJ., Sacroiliac luxation in the cat: defining a safe corridor in the dorsoventral plane for screw insertion in lag fashion, „Vet Surg.” 2009.